Abstract

The pressed powder sample was determined by Epsilon 5 high-energy polarized energy dispersive X-ray fluorescence (EDXRF) spectrometer. The correlation coefficient of the standard curve method is 0.991 0. The precision is 2.98% and the relative deviation of the measurement standard samples is up to 6.40%, which showed that the precision and accuracy of the method were also good. Simultaneous measurement of seven soil samples with this method and γ-spectrometer were carried, the results of two analytical methods was compared using a paired t-test by SPSS program, which showed that there was no significant difference in the two sets of data. The results predicated that EDXRF could be a potential simple method for analyzing 40K in soil samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.