Abstract

A rapid and convenient method for the determination of isoniazid in blood serum or plasma has been described. This method is based on the reaction between INH and sodium pentacyanoammineferroate, which results in a yellowish chromogen whose maximum absorption is at 430 mμ. Factors that influence the reaction are discussed in order to attain the maximal sensitivity. In comparison with other methods for INH estimation, the present method is superior, since it is specific and simple. It can be carried out in not more than half an hour with ordinary laboratory equipment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.