Abstract
A rapid and inexpensive micro-assay for determining cholesterol in plasma and isolated lipoprotein fractions has been established which utilizes a commercially available enzymatic reagent with semi-automated instruments and microtiter plates. The assay is sensitive, precise, and easy to perform. The color development is linear from 0.4 to 20 micrograms cholesterol/well, with sample volumes of 2 to 100 microliters. Inter- and intra-assay variability yielded coefficients of variation (CV) of 2.75% (n = 51) and 1.09% (n = 32), respectively. The concentrations of total plasma and lipoprotein cholesterol (d greater than 1.006 g/ml) obtained with this method were compared with those analyzed in a lipid laboratory standardized to the Centers for Disease Control. The correlation coefficients between the two methods were 0.976 and 0.964, respectively. For total high density lipoprotein (HDL) and the HDL3 subfraction, inter-assay variability was 4.12% and 6.33% (n = 27), respectively; the intra-assay variability was 2.79% and 4.19% (n = 12).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.