Abstract

In this novel enzyme-tagged immuoelectrochemical assay, disposable carbon felt discs serve both as electrodes and as the heterogeneous solid phase. Antibodies are immobilized on the carbon felt via a diaminoalkane-biotin-avidin-biotin bridge. Alkaline phosphatase is used as a label. Bound antibodies are monitored by following the electro-oxidation of aminophenol, produced enzymatically from p-amino-phenyl phosphate by the immobilized alkaline phosphatase at the electrode surface. A model system designed for determination of mouse IgG concentration yielded a calibration curve ranging from 10 pg/ml to 100 micrograms/ml. This assay can be performed rapidly and a single determination completed within 20 minutes. The system is useful also for rapid quantitation of a small number (approximately 80 organisms per ml) of bacteria.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.