Abstract

In cyanotoxin measurements, effective release of intracellular cyanotoxins through cell lysis is pivotal. The conventional method for cell lysis is repeated freeze-thaw (F-T), which has several disadvantages, including poor reproducibility since it is operator and equipment dependency and time-consuming. In this study, a rapid and sensitive method was developed using irreversible electroporation, reducing quantification time by over 6 h compared to F-T. Focusing on microcystins (MCs), we developed the most optimal electroporation medium (50 mM Tris (pH 7.0) with 0.5 % SDS) and determined the optimal intensity of electroporation using Microcystis culture. Microcystis cell rupture was validated by scanning electron microscopy. COMSOL simulations mirrored experimental conditions. Compared to F-T, this new method generated an average 13.7 % (6.7 ppb) more MCs from lake water samples (p ≥ 0.05). This innovation, surpassing the time-consuming F-T process, emerges as a valuable tool for timely decision-making in water safety advisory and cyanotoxin management in various settings.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.