Abstract

At one blood center, each of 20 donors underwent plateletpheresis on four blood cell separators in random order. We compared the CS3000+, Amicus V 2.41, MCS Plus, and Spectra LRS V 7 Turbo regarding platelet (PLT) yield, pre- and post-procedure PLT counts, percent fall in donor PLT count, process time, efficiency, PLT product and donor PLT volume (MPV). Using >or= 150 x 10(9) PLTs/L pre-donation counts, a goal was set of 4.5 x 10(11) PLTs unit in up to 100 minutes processing time. Results were (mean values) PLT yields of Amicus, Spectra, CS3000+, and MCS Plus: 4.3, 4.6, 4.3, 4.0 x 10(11) PLTS, respectively; percent donor PLT fall: 24, 32, 30, 29%, respectively; processing times: 50, 74, 87, 101 minutes, respectively; relative efficiency (RE): 2.2, 1.6, 1.2,1.0, respectively (based on the MCS Plus performance with RE of 1 = 4 x 10(9) PLTS/min); PLT product MPV: 6.7, 7.4, 6.8,7.1 fL, respectively; pre-procedure donor MPV: 7.7, 7.3, 7.6 and 7.6 fL, respectively; and percent donor MPV change: -5.2, 0, -6.6, and -10%, respectively. Significant changes in the donor MPV were noted (P < 0.05) but could not be related to product MPV. Spectra seemed to collect larger PLTs (higher MPV); the significance remains unknown for both donors and recipients. Importantly, all four separators gave acceptable and comparable PLT yields (P < 0.05) with Spectra trending higher. The short process time and high RE together indicate highly efficient collections particularly by Amicus and Spectra.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call