Abstract

We compared insertion rates of single-use polyvinyl chloride laryngeal mask airways (LMAs) vs single-use silicone LMAs in 72 anaesthetised patients. Both airways were produced by Flexicare Medical. Laryngeal mask airway insertion was successful on the first attempt in 68/72 (94%) polyvinyl chloride LMAs vs 64/72 (89%) silicone LMAs (p = 0.39). Overall insertion rates were 72/72 (100%) for the polyvinyl chloride LMAs and 71/72 (99%) for the silicone LMAs (p = 1.0). Mean (SD) insertion times were similar for polyvinyl chloride and silicone LMAs: 24.3 (5.1)s vs 24.8 (7.8)s (p = 0.64). Laryngeal mask airway position, as assessed using a fibrescope, was not different (p = 0.077). The median (IQR [range]) leak pressure was 16 (12-20 [6-30]) cmH(2) O for the polyvinyl LMA and 18 (13-22 [6-30]) cmH(2) O or the silicone LMA (p = 0.037). In conclusion, we did not find any important differences between polyvinyl chloride and silicone laryngeal mask airways.