Abstract
This study assessed the reliability of radiological assessment of polyethylene wear in loose total hip replacements (THRs) by comparing it with a laboratory assessment using the shadowgraph technique. It was performed in Avon Orthopaedic Centre, Southmead Hospital, from September 1997 to June 1998. True linear wear depth measured using the shadowgraph technique was 42% greater than that determined from the radiological femoral head eccentricity in standard non-weightbearing plain x-rays. There was an inverse statistically significant correlation between cup inclination angle and the volumetric wear rate. The linear wear rate, as determined with the shadowgraph technique, was greater in retrieved cups during the first six years postoperatively, but reduced to a lower level in cups retrieved after that time of service.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.