Abstract

In this study, a method of determining radiochemical yield and radiochemical purity using radio-HPLC detection employing a dual-flow-cell system is evaluated. The dual-flow cell, consisting of a reference cell and an analytical cell, was constructed from two PEEK capillary coils to fit into the well of a NaI(Tl) detector. The radio-HPLC flow was directed from the injector to the reference cell allowing on-line detection of the total injected sample activity prior to entering the HPLC column. The radioactivity eluted from the column was then detected in the analytical cell. In this way, the sample will act as its own standard, a feature enabling on-line quantification of the processed radioactivity passing through the system. All data were acquired on-line via an analog signal from a rate meter using chromatographic software. The radiochemical yield and recovery could be simply and accurately determined by integration of the peak areas in the chromatogram obtained from the reference and analytical cells using an experimentally determined volume factor to correct for the effect of different cell volumes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.