Abstract

A simple five parameter model for the efficiency of a Si(Li) detector has been developed. It was found necessary to include a radially dependent efficiency even for small detectors. The model is an extension of the pioneering work of Hansen et al. [1] but correction factors include more up to date data and explicit equations for the mass attenuation coefficients over a wide range of photons energies. Four of the five parameters needed are generally supplied by most commercial manufacturers of Si(Li) detectors. 54Mn and 241Am sources have been used to calibrate a Si(Li) to ∼±3% over the energy range 3–60 keV.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.