Abstract

A description is presented of a simple and sensitive method for detecting a change in the energy of the electrons bombarding the target of medical accelerators. This technique is useful for x-ray beams with end point energy in the range of 15.7 to 25 MeV. The method is based on the photoactivation of 16O and 14N in a small sample of ammonium nitrate. It was found that the ratio of the activity induced in the oxygen divided by that produced in the nitrogen can be used as a quality control technique to detect a change in the energy of the electrons that bombard the target of the accelerator. An electron energy change of the order of 0.2 MeV can be determined using this method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.