Abstract

Background 3D and gapless multi-slice 2D cine Phase contrast (PC) MRI enables the visualization and analysis of 4D (3D + time) intracardiac flow. However, the image quality which affects the accuracy of segmentation of left ventricle (LV) has not been discussed. In this study, we compared the magnitude image and the source image of PC MR angiography (MagSUM) derived from both 2D and 3D PC MRI. For the qualitative analysis, we assessed the image quality for the delineation of endocardial border. And we measured LV volume and compared with standard cine MRI for the quantitative analysis. Methods Ten healthy volunteer (all male, median age: 24 years) underwent 2D and 3D prospectively ECG-gated PC MRI with respiratory naviation on 3T MRI (Trio a Tim, Siemens AG Healthcare Sector, Erlangen, Germany).The short axial plane was employed to cover left ventricle. The common parameters for PC MRI were: VENC = 150 cm/ sec, segmentation factor = 2 to 3, reconstruction phase = 12 to 14 according to R-R interval. The acquisition voxel of 2D and 3D PC MRI were reconstructed into 2.1x2.1x5 mm and 2.1x2.1x2.1 mm, respectively. Standard 2D retrospectively ECG-gated cine MRI was reconstructed to 2.1x1.8x8 mm. Two readers discussed and determined the

Highlights

  • 3D and gapless multi-slice 2D cine Phase contrast (PC) MRI enables the visualization and analysis of 4D (3D + time) intracardiac flow

  • The image quality which affects the accuracy of segmentation of left ventricle (LV) has not been discussed

  • The acquisition voxel of 2D and 3D PC MRI were reconstructed into 2.1x2.1x5 mm and 2.1x2.1x2.1 mm, respectively

Read more

Summary

Open Access

A qualitative and quantitative assessment of cardiac cine Phase contrast MRI: comparison of image quality between 2D and 3D acquisition. Munemura Suzuki1,2*, Takahiko Nakazono, Ken Yamaguchi, Kiyoshi Dogomori, Yoshiaki Komori, Norihiko Kotooka, Koichi Node, Hiroyuki Irie. Background 3D and gapless multi-slice 2D cine Phase contrast (PC) MRI enables the visualization and analysis of 4D (3D + time) intracardiac flow. The image quality which affects the accuracy of segmentation of left ventricle (LV) has not been discussed. We compared the magnitude image and the source image of PC MR angiography (MagSUM) derived from both 2D and 3D PC MRI. We assessed the image quality for the delineation of endocardial border. We measured LV volume and compared with standard cine MRI for the quantitative analysis

Methods
Results
Conclusions

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.