Abstract

Introduction: To develop and characterize the accuracy and reproducibility of a 'quad-phantom' dosimeter which will serve as an independent verification tool during commissioning of a PRESAGE/optical-CT 3D dosimetry system. Methods: A 16cm × 12cm cylindrical quad-phantom was constructed from four pieces of solid polyurethane mimicking the PRESAGE material. Films were placed and anchored in orthogonal planes and the quad-phantom was fastened tightly together and placed in a water-filled Styrofoam container for irradiation. A simple, two-field plan consisting of 6×6cm anterior-posterior and right-lateral 6MV photon beams (400cGy) was delivered three times (fresh films inserted for each) with a Varian Clinac 600C. Image registration was performed in the Computational Environment for Radiological Research (CERR) and dose profiles and gamma analysis was performed in CERR and MATLAB. Results & Discussion: Excellent reproducibility was observed during the irradiations, with ∼2.3% standard deviation between all pixels. Using a 3%, 3mm gamma criteria, excellent dosimetric accuracy was observed, with 98.8% and 96.3% passing rates in the sagittal and axial planes, respectively. Conclusion: The preliminary results indicate that the quad-phantom can serve as a reproducible and accurate system for high resolution dosimetry in orthogonal planes and should serve as an effective verification tool for PRESAGE/optical-CT in more challenging clinical scenarios.