Abstract
A neutron generator system utilizing two Zetatron neutron tubes has been designed and delivered to the U.S. Department of Agriculture Human Nutrition Research Center on Aging at Tufts University for use in clinical measurements of body carbon by neutron inelastic scattering. Each neutron tube is capable of delivering 10 3−10 4 14-MeV neutrons in a 7-μs pulse at repetition rates of 4 or 8 kHz, and can be operated independently as well as in a master-slave mode. The neutron tubes are gas filled with a mixture of deuterium and tritium; the target of the tube is operated at − 30 to − 60 kV dc and the ion source is operated with a 2.5-kV, 7-μs pulse. The tube gas pressure is monitored and controlled by measuring the total current in the high voltage circuit and feeding it back to the gas-reservoir drive circuit. Neutrons were measured with a plastic scintillator and photomultiplier tube.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.