Abstract
Negative pressure is a problem in pulsatile cardiopulmonary bypass (CPB). To avoid this, the authors designed a pulsatile CPB system containing a Sarns centrifugal pump (CP) and a Univox membrane oxygenator, in which the inertial flow is not obstructed by the CP. In both an in vitro study and a clinical study, negative pressure was not observed in the arterial line of the CPB circuit when this system was used. When a roller pump (RP) was used, however, instead of a CP, negative pressure did occur. In a clinical study using this system, mean pulse pressure was 36 mmHg and hemolysis, expressed as the rate of rise in plasma free hemoglobin from 10 to 70 min of CPB, was 26.2 mg/dl/hr, which did not exceed that seen with a pulsatile CPB using an RP instead of a CP. The hemolysis seen in the study caused no clinical problems. Thus, pulsatile CPB using a CP and Univox membrane oxygenator should be considered for clinical use to prevent the occurrence of negative pressure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.