Abstract The sensitivity of Neutron Macromolecular Crystallography to the presence of hydrogen makes it a powerful tool to complement X-ray crystallographic studies using protein crystals. The power of this technique is currently limited by the relative low neutron flux provided by even the most powerful neutron sources . The strong polarization dependence of the neutron scattering cross section of hydrogen will allow us to use Dynamic Nuclear Polarization to dramatically improve the signal to noise ratio of neutron diffraction data, delivering order of magnitude gains in performance, and enabling measurements of radically smaller crystals of larger protein systems than are possible today. We present a prototype frozen spin system, built at Oak Ridge National Laboratory to polarize single protein crystals on the IMAGINE beamline at the High Flux Isotope Reactor (HFIR). Details of the design and construction will be described, as will the performance of the system offline and during preliminary tests at HFIR.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.