Abstract

The aim of the present study was to determine the accuracy of breast cancer measurement with 2 D shear wave elastography (2 D SWE), B-mode ultrasound, and mammography by comparing these methods with the actual histopathological tumor size. 135 patients with proven malignant breast lesions at two centers were included in a prospective study. The maximum lesion diameter was measured by 2 D SWE, B-mode ultrasound, and mammography. After surgery, the histopathological tumor size was measured by a pathologist. The mean difference in tumor size on 2 D SWE, B-mode ultrasound, and mammography on the one hand, and the actual tumor size on the other, was -0.03 cm, 0.16 cm and 0.10 cm, respectively. Lesion size on histopathological investigation was overestimated by 2 D SWE (p = 0.004) and underestimated by B-mode imaging (p < 0.001). All three imaging methods underestimated the size of invasive lobular cancers and lesions > 15 mm; 2 D SWE was most accurate in this regard. 2 D SWE predicted lesion size more precisely than B-mode ultrasound or mammography. In cases of invasive lobular carcinoma, all three imaging methods underestimated lesion size, with 2 D SWE coming closest to the actual tumor size.