Abstract

The purpose of this study was to compare the accuracy and safety of an instrumented posterior lumbar interbody fusion (PLIF) using a robot-assisted minimally invasive (Robot-PLIF) or a conventional open approach (Freehand-PLIF). Patients undergoing an instrumented PLIF were randomly assigned to be treated using a Robot-PLIF (37 patients) and a Freehand-PLIF (41 patients). For intrapedicular accuracy, there was no significant difference between the groups (P = 0.534). For proximal facet joint accuracy, none of the 74 screws in the Robot-PLIF group violated the proximal facet joint, while 13 of 82 in the Freehand-PLIF group violated the proximal facet joint (P < 0.001). The average distance of the screws from the facets was 5.2 ± 2.1 mm and 2.7 ± 1.6 mm in the Robot-PLIF and Freehand-PLIF groups, respectively (P < 0.001). Robotic-assisted pedicle screw placement was associated with fewer proximal facet joint violations and better convergence orientations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.