Abstract
Modifications in the design of knee replacements have been proposed in order to maximise flexion. We performed a prospective double-blind randomised controlled trial to compare the functional outcome, including maximum knee flexion, in patients receiving either a standard or a high flexion version of the NexGen legacy posterior stabilised total knee replacement. A total of 56 patients, half of whom received each design, were assessed pre-operatively and at one year after operation using knee scores and analysis of range of movement using electrogoniometry. For both implant designs there was a significant improvement in the function component of the knee scores (p < 0.001) and the maximum range of flexion when walking on the level, ascending and descending a slope or stairs (all p < 0.001), squatting (p = 0.020) and stepping into a bath (p = 0.024). There was no significant difference in outcome, including the maximum knee flexion, between patients receiving the standard and high flexion designs of this implant.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The Journal of Bone and Joint Surgery. British volume
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.