Abstract

Background: Persistent use of real-time continuous glucose monitoring (CGM) improves diabetes control in individuals with type 1 diabetes (T1D) and type 2 diabetes (T2D).Methods: PRECISE II was a nonrandomized, blinded, prospective, single-arm, multicenter study that evaluated the accuracy and safety of the implantable Eversense CGM system among adult participants with T1D and T2D (NCT02647905). The primary endpoint was the mean absolute relative difference (MARD) between paired Eversense and Yellow Springs Instrument (YSI) reference measurements through 90 days postinsertion for reference glucose values from 40 to 400 mg/dL. Additional endpoints included Clarke Error Grid analysis and sensor longevity. The primary safety endpoint was the incidence of device-related or sensor insertion/removal procedure-related serious adverse events (SAEs) through 90 days postinsertion.Results: Ninety participants received the CGM system. The overall MARD value against reference glucose values was 8.8% (95% confidence interval: 8.1%–9.3%), which was significantly lower than the prespecified 20% performance goal for accuracy (P < 0.0001). Ninety-three percent of CGM values were within 20/20% of reference values over the total glucose range of 40–400 mg/dL. Clarke Error Grid analysis showed 99.3% of samples in the clinically acceptable error zones A (92.8%) and B (6.5%). Ninety-one percent of sensors were functional through day 90. One related SAE (1.1%) occurred during the study for removal of a sensor.Conclusions: The PRECISE II trial demonstrated that the Eversense CGM system provided accurate glucose readings through the intended 90-day sensor life with a favorable safety profile.

Highlights

  • Numerous studies have shown that use of real-time continuous glucose monitoring (CGM) improves glycemic control[1,2,3,4,5,6] and quality of life[7] in individuals with type 1 diabetes (T1D) and type 2 diabetes (T2D).[8]

  • The overall mean absolute relative difference (MARD) value against reference glucose values was 8.8% (95% confidence interval: 8.1%–9.3%), which was significantly lower than the prespecified 20% performance goal for accuracy (P < 0.0001)

  • The PRECISE II trial demonstrated that the Eversense CGM system provided accurate glucose readings through the intended 90-day sensor life with a favorable safety profile

Read more

Summary

Introduction

Numerous studies have shown that use of real-time continuous glucose monitoring (CGM) improves glycemic control[1,2,3,4,5,6] and quality of life[7] in individuals with type 1 diabetes (T1D) and type 2 diabetes (T2D).[8] The clinical benefits of CGM technology are maximized in patients who regularly wear their CGM devices for at least 6 days per week[1,2,3,5,6,9,10]; many patients struggle to achieve consistent adherence.[9,10] many individuals who a Mark P. The primary safety endpoint was the incidence of device-related or sensor insertion/removal procedure-related serious adverse events (SAEs) through 90 days postinsertion. Conclusions: The PRECISE II trial demonstrated that the Eversense CGM system provided accurate glucose readings through the intended 90-day sensor life with a favorable safety profile

Methods
Results
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call