Abstract
BackgroundSequential compression devices (SCDs) reduce deep venous thrombosis in postsurgical patients, but the use is hindered by poor compliance. MethodsGeneral and orthopedic surgery patients (n = 67) were randomized to standard- or battery-powered SCDs. Compliance was documented hourly. Nurses and patients were issued a survey to assess barriers to compliance and device satisfaction. ResultsCompliance with standard SCDs was 47% compared with 85% with battery-powered SCDs (P < .001). The most common barriers identified by nurses and patients were ambulation and transfers, which were mitigated with the battery-powered device. A majority (79%) of those issued a battery-powered device reported no major problems compared with only 14% of patients issued a standard device (P < .005). ConclusionsThe dual venous thromboembolism prevention strategies of early mobilization and SCD utilization can be met with the appropriate equipment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.