Abstract

Purpose: The study compared the setup reliability of 3 patient immobilization systems, a rubber leg cushion, the alpha cradle, and the thermoplastic Hipfix device, in 77 patients with cT1–T3, N0, M0 prostate cancer receiving conformal radiotherapy. Methods and Materials: Port films were analyzed and compared to simulation films to estimate the setup errors in the three coordinate axes (anterior–posterior, cranial–caudal, medial–lateral). A total vector error was calculated from these shifts. Results: The Hipfix was found significantly superior to the other two devices in reducing mean setup errors in all axes ( p < 0.005). The average field-positioning error with the Hipfix ranged from 1.9 mm to 2.6 mm for all axes, whereas the deviation for the other two systems ranged from 2.7 to 3.4 mm. Errors greater than 10 mm were virtually eliminated with the Hipfix system. There was a reduction in the mean total vector error in the alpha cradle and Hipfix patient cohorts over time, reflecting improved efficacy as a result of experience. Conclusion: There was a significant difference in the performance of each immobilization device. The Hipfix was consistently more reliable in reducing setup errors than the other devices.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.