Abstract
ObjectivesTo evaluate the performance of dynamic contrast-enhanced MRI measurement of glomerular filtration rate (GFR) compared with the reference standard technique of urinary clearance of 51Cr-EDTA. Patients and methodsAll kidney transplant recipients (KTRs) with an indication for non-urgent contrast-enhanced MRI at our institution were prospectively included between 2008 and 2012. Renographies were acquired by low-dose dynamic contrast-enhanced MRI (DCE-MRI) then fitted with a two-compartment pharmacokinetic model. MR-GFR was compared with reference isotopic measurements using Bland-Altman diagrams, intraclass correlation coefficient (ICC) and concordance rates. ResultsForty-two KTRs (mean age 51.5 years, 26–74) were analyzed. Mean estimated GFR was 48.5 ± 27 mL/min/1.73m2 (24–178 mL/min). The mean bias was +13.2 mL/min (6.4–20.0, +36.9%) ranging from -31.0 mL/min (-41.7%) to +101.4 mL/min (+89.2%) with a large variability (standard-deviation: 22.3 mL/min; limits of agreement: [-30.6 (-43.3–-18.9); +57.0 (45.3–68.7)]). The ICC was 0.32 (0.02–0.56) and the concordance rate was 28.6% (14.9–42.2). ConclusionsThe large variability of MR-GFR compared with the reference technique precludes its use in KTRs, whose anatomical peculiarities make standardization of arterial input function (AIF) difficult.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.