Abstract

Post-endoscopic submucosal dissection coagulation syndrome (PECS) is a recognized complication of colorectal endoscopic submucosal dissection (ESD); however, there is a lack of interventions for preventing PECS. We therefore conducted a prospective study to evaluate the utility of maXium, a novel electrosurgical unit, for preventing PECS. This single-center, prospective cohort study prospectively enrolled patients undergoing colorectal ESD. The voltage and power of the electrosurgical units were measured. PECS was defined as a visual analog scale (VAS)≥30mm, an increase of VAS≥20mm from baseline, body temperature ≥37.5°C, or white blood cell count ≥10000/μL after ESD. PECS was classified into type I (without extra-luminal air) and type II (with peri-luminal air). The primary endpoint was the incidence of PECS. A sample size of 92 patients was required to ensure the upper limit of the 90% CI for the incidence of PECS was less than 15%. At resistances greater than 400Ω, the maXium unit allowed submucosal dissection with lower power than with the VIO300D unit. Ninety-one patients meeting the inclusion criteria were included in the final study analysis. The incidence of PECS was 16% (90% CI, 10-23%), comprising type I (11%) and type II (5%) PECS. Simple extra-luminal air without PECS was observed in 7% of patients. Use of the maXium electrosurgical unit did not reduce the incidence of PECS after colorectal ESD; however, the maXium unit had equivalent performance to a conventional electrosurgical unit used for colorectal ESD.