Abstract

Surgical instrument trays are wrapped for sterilization, and these wraps are inspected by operating room personnel for defects before using the instruments. In this study, we intentionally damaged wraps, and these wraps were evaluated by operating room personnel. Examiners correctly identified a wrapper's sterile integrity with an overall 56.1% accuracy, with correct identification of 67.2% for defects between 2 and 5 cm. However, studies have shown bacterial contamination through defects as small as 1.1 mm. This study suggests that the current method for assessing sterility is inadequate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.