Abstract
PurposeA number of guidelines for ultrasound quality assurance recommend the use of the in air reverberation depth as a proxy measure for sensitivity. The test is quantised, i.e. it depends on the brightness of the deepest in air reverberation. The aim of this study was to investigate a possible enhancement to the test, where the gain is reduced to determine the “reverberation threshold”. MethodsThe test was introduced in several ultrasound departments. Results were audited to determine agreement with annual tests of sensitivity using a tissue mimicking test object. ResultsThe new test was performed on 100 probes. A change in reverberation threshold was demonstrated in 9 probes; 8 of these also had changes in penetration and/or grey level in a tissue mimicking test object. Reduced penetration but no change in reverberation threshold was seen in 2 probes. ConclusionsThe reverberation threshold provides a simple enhancement to the in air sensitivity test. Periodic sensitivity testing with a tissue mimicking test object remains important.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.