Abstract

Ischemia-reperfusion injury remains a fundamental problem during organ transplantation logistics. One key technical factor is the rapid allograft rewarming during the time of vascular reconstruction in the recipient. In this pilot study, a new thermal insulation bag (TIB) for organ transplantation was used. Insulation capacity, tissue compatibility, and usability were tested initially ex vivo on porcine kidneys (n = 24) followed by the first in vivo usage. Fourteen female German landrace pigs underwent kidney auto-transplantation after 24 h cold storage (4 °C). During the implantation process the kidney was either insulated with the new TIB, or it was not thermo-protected at all, which represents the clinical standard. In this proof-of-concept study, the usability (knife-to-skin-time) and the general thermal capacity (30 min warm storage at 38 °C ex vivo p < 0.001) was shown. The clinical outcome showed significant differences in the determination of CRP and pi-GST levels. Syndecan-1 Antibody staining showed clear significant higher counts in the control group (p < 0.01) indicating epithelial damage. However, the effect on renal outcomes in not severely pre-damaged kidneys does not appear to be conclusively significant. A close follow-up study is warranted, especially in the context of marginal organs or in cases where anastomosis-times are prolonged due to surgical complexity (e.g., multiple vessels and complex reconstructions).