Abstract

We present a method based on pre- and postscanning a piezoelectric tube scanner used in a force probe that improves the reproducibility of the scan lengths. Instead of prescanning in the same direction as when acquiring data (the z direction), which could destroy a sensitive surface, we perform lateral (x/y direction) prescans. As lateral motions of the tube scanner involve out of phase elongations and compressions of the tube in the z direction, these kinds of prescans will have a stabilizing effect on the z motion as well. By adding an additional postscan in the ±z directions, we reduce the piezoelectric creep following the data acquisition scan. When comparing the lengths of z scans with and without the pre/postscan procedure, preceded by a z voltage step 60 s before data acquisition, the deviation between four consecutive scans improved from 12% to 1.4%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.