Abstract

Data are currently being collected from hospital radiology information systems in the North West of the UK for the purposes of both clinical audit and patient dose audit. Could these data also be used to satisfy quality assurance (QA) requirements according to UK guidance? From 2008 to 2009, 731 653 records were submitted from 8 hospitals from the North West England. For automatic exposure control QA, the protocol from Institute of Physics and Engineering in Medicine (IPEM) report 91 recommends that milliamperes per second can be monitored for repeatability and reproducibility using a suitable phantom, at 70-81 kV. Abdomen AP and chest PA examinations were analysed to find the most common kilovoltage used with these records then used to plot average monthly milliamperes per second with time. IPEM report 91 also recommends that a range of commonly used clinical settings is used to check output reproducibility and repeatability. For each tube, the dose area product values were plotted over time for two most common exposure factor sets. Results show that it is possible to do performance checks of AEC systems; however more work is required to be able to monitor tube output performance. Procedurally, the management system requires work and the benefits to the workflow would need to be demonstrated.