Abstract
To determine the microbial content and endotoxin concentration in the water reservoirs of benchtop autoclaves used in general dental practice. The study was done in two stages. Firstly water samples were taken daily from the reservoirs of 20 autoclaves used in general dental practices for six days. The microbial content and endotoxin concentration was determined. Second the practitioners were instructed to wash and clean the reservoirs of the autoclaves each morning prior to refilling with fresh water. The reservoirs were sampled in the evenings after a day's use and the microbial content and endotoxin concentration determined. The total viable count of bacteria before cleaning varied from 2,300 - 8 x 10(4) cfu/ml and after from 0-40 cfu/ml. The endotoxin concentrations before cleaning varied from 360-2,200 EU/ml and after 0-80 EU/ml. The reservoirs of non-vacuum benchtop autoclaves can become severely contaminated with micro-organisms particularly Gram-negative bacteria. The endotoxin content of the reservoirs can get to high and unacceptable concentrations. Draining the water reservoir and cleaning on a daily basis reduces the microbial content and the endotoxin concentration to acceptable levels.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.