Abstract

An account is given of a series of precision experiments to investigate cut-off effects in a parallel-plate type of free-air chamber, and the relationship of these effects to the chamber geometry is analysed. The discussion is extended to enable estimates to be made of the errors associated with other parallel-plate chambers. In conclusion, it is shown that deficiencies in the geometry of the standard chamber in use at present at the National Physical Laboratory in England have led to errors approaching 3 per cent in the realisation of the röntgen at ordinary deep-therapy qualities. The corresponding error in the National Bureau of Standards chamber in the United States is estimated to lie between 1 and 2 per cent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.