Abstract
We have remodeled the X-ray spectra estimation method, originally proposed by Waggener et al. (Med. Phys. 26(1999)1269), based on the iterative perturbation technique to minimize differences between measured and calculated transmission curves using low Z attenuators. With our approach, the iterative perturbation cycle is repeated until a reasonable spectrum is obtained. Furthermore, the degree of differences between the measured and calculated transmission curves is also checked using high Z attenuators. The present experimental study, conducted using 4, 10 and 15 MV X-ray beams from a linear accelerator, demonstrated that the spectrum varies strongly with the off-axis distance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.