Abstract

A 12 mm tissue-equivalent proportional counter (TEPC) was designed, constructed and successfully tested. This detector achieves features common to others TEPCs, but it does not use field shaping electrodes and it is able to work at higher bias voltages which makes it capable of measuring the whole range of energy deposition events for gamma rays. The following approach was used to design the detector: first of all the use of a cylindrical shape detector featured as simple as possible but keeping its performance as well, the next point is to avoid the use of field shaping tubes and the last one is to make the preamplifier small and as close as possible to the detector. Its construction is based on a cylindrical proportional counter with A-150 tissue-equivalent material as cathode, TE-methane gas as a proportional gas and a 20 microns diameter wire as anode.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call