Abstract
A potentiometric acetylcholinesterase biosensor based on pH-sensitive PVC membrane with plasma-polymerized ethylenediamine film has been proposed. Glow discharge plasma technique was utilized to deposit a film containing amino groups on the surface of pH-sensitive PVC membrane. Such an approach makes it possible to fabricate a very simple self-mounted acetylcholinesterase membrane directly attached to PVC pH-sensing electrode surface. The deposition conditions were optimized and the optimum plasma polymerization parameters were as follows: applied power, 75 W; pressure, 130 Pa; plasma exposure time, 60 min and flow rates, 10 ml min −1. Two neutral carriers, DOODA and DAMAB were compared as the pH-carrier, and the biosensor based on DAMAB showed better response characteristics. The effect of pH of the medium was also examined. The potential responses of biosensor increase with the acetylcholine chloride concentration over the concentration range of 10 −6 to 10 −1 mol dm −3 and the detection limit is 0.002 mmol dm −3 in pH 8.0 PB buffer solution. The biosensor was employed more than 200 times in 20 days while the maximum response span of biosensor retained 90%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.