Abstract

The performance of a position sensitive gamma-ray scintillator detector (PSD) is described. This PSD is based on a lutetium yttrium oxyorthosilicate (LYSO) crystal read out by a crossed-wire anode position sensitive photomultiplier tube (PSPMT). The main difference with respect to similar existing devices is the individual multi-anode readout (IMAR) approach that is followed here. This method allows to exploit better the intrinsic characteristics of the PSPMT, thus yielding better linearity, improved spatial resolution, and a larger field of view. The new detector is intended for the characterization of 3-D position sensitive germanium detectors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.