Abstract

As patients with left ventricular (LVAD) and biventricular assist devices are supported for increasingly long durations while awaiting heart transplantation or cardiac recovery, there is a need to facilitate greater patient mobility and ambulation. To meet these needs, the authors have developed the Thoratec TLC-II Portable VAD Driver, which is a small brief-case sized (33 x 34 x 13 cm) pneumatic unit for Thoratec's paracorporeal and implantable VADs. The TLC-II consists of an electric motor driven air compressor for supplying both positive and negative air pressure, solenoid valves for switching between LVAD/RVAD filling and ejection, and microcontroller based electronics and firmware. Four power sources are provided: external power, two rechargeable lithium-ion battery packs, and an emergency battery that drives an independent electronic back-up system. The 8 kg TLC-II can be carried by hand, with a shoulder strap, or pushed on a wheeled mobility cart. Trend information stored in the TLC-II can be accessed by an external system computer with a color touchscreen mounted on a docking station, which also houses the battery charger. Control configurations (univentricular/biventricular operation, beat rates, etc.) are entered on the touchscreen and programmed into the TLC-II. In vitro testing demonstrates the ability to pump VAD outputs up to 7 L/min. By providing improved patient mobility, this small driver will enhance rehabilitation and improve the quality of life of VAD patients.