Abstract
The results of the development and research of the characteristics of a portable γ-spectrometer based on a segmented high-purity Ge detector with background suppression and a detection efficiency of 60% are presented. In comparison to the conventional detection of radiation, when the entire detector volume is enabled, the relative improvement of the peak-to-Compton ratio for energies of 662 and 136 keV amounts to a factor of 1.4 and above 1.5, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.