Abstract

A point-focusing small angle x-ray scattering (SAXS) camera using a doubly curved monochromator of a W/Si multilayer has been designed, constructed, and tested. The two radii of curvature of the monochromator are 20 400 and 7.6 mm. The reflectivity of its first-order Bragg reflection for CuKα radiation was calculated to be 0.82, being comparable to that (0.81) of its total reflection. By only 10 s x-ray exposure, scattering from a high-density polyethylene film was detected on an imaging plate (IP). A rotating-anode x-ray generator operated at 40 kV and 30 mA was used. Diffraction from rat-tail collagen has shown that the optical arrangement gives the Bragg spacing up to, at least, 30 nm for CuKα radiation. Combined with IPs, the camera may permit us to carry out time-resolved SAXS measurements for phase behaviors of liquid crystals, lipids, polymer alloys, etc., on conventional x-ray generators available in laboratories.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.