Abstract

This exercise focused on performance of NBS quality monitoring establishments with respect to enumeration of low leucocyte and other quality indexes of platelet concentration. Paired identical leucodepleted platelet samples, spiked with WBC ( 20 cells/ μl ) in `vacuette' or `pouch' were assessed by participants ( n=20) on days 1, 2 and 5. For low WBC counting, all laboratories gave estimates within acceptable range (±25%) and good agreement between storage and assay methods was observed on days 1 and 2. Day 5 results showed greater variability. Under improved performance criteria (±15%), only one laboratory under-estimated at days 1 and 2. Similarly, other parameters demonstrated good agreement between storage methods on days 1 and 2. At day 5, mean results were often significantly different to previous days. Improved performance target (±15%) will allow identification of non-conformers.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.