Abstract
A first order electrostatic ion mirror has been constructed and mounted on a time-of-flight spectrometer coupled to the Uppsala EN-tandem accelerator at The Svedberg Laboratory. In electronic sputtering of CsBr with MeV 127I ions a highest mass resolution around 10500 at a full-width at half maximum for the cluster ions (CsBr) 2Cs + has been obtained. The resolution for organic molecules is somewhat lower than this value but still significantly better than in a linear system. The mirror also provides great simplification of the calibration procedure and an improvement in mass accuracy compared with a linear spectrometer. For stable ions, and with the help of an einzel lens, the transmission approaches 100% in the mirror. The mirror is equipped with two stop detectors which make it possible to study the decay of metastable ions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Mass Spectrometry and Ion Processes
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.