Abstract

The resolving function and the shape of the rocking curves are calculated for plane-crystal neutron spectrometers,taking into account the effect of total reflection from the collimator walk. The features of crystal-type neutronspectrometers are described for the plane-crystal and bent-crystal variants, and results obtained with such a spectrometer are presented. When using a calcite crystal, the resolving power of the instrument is 0.53 μsec/m. In this case, the energy range extends up to 5.5 ev. Results obtained with a bent-quartz crystal are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.