Abstract
AbstractThis paper presents a planar microwave sensor for the noninvasive detection of glucose concentration in diabetic patients. The designed sensor operates from the 3.8 to 6.2 GHz frequency band, which covers the 5.8 GHz Industrial Scientific and Medical (ISM) band. The designed sensor shows a percentage bandwidth of 23.8% with a reflection coefficient (S11) of −50 dB at the resonance frequency of 5.7 GHz. The detection was carried out by varying the relative permittivity of the blood in accordance with the glucose concentration based on the Cole–Cole model. The measured result is calculated in terms of varying resonance frequency with variation in the reflection coefficient |S11| of the designed sensor. The observed frequency shift and corresponding sensitivity of the sensor are found at 1.7 GHz and 0.089 MHz/mg dL−1, respectively. An experimental validation has also been performed, and the frequency shift is analyzed by interacting the human thumb with the sensor. The simulated and experimental results of the designed sensor suggest that it can be useful for detecting glucose concentration noninvasively for diabetic patients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Microwave and Wireless Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.