A pilot study evaluating the use in paediatric polytrauma of the STATSCAN, a low-radiation dose, fan-beam digital radiography unit (Lodox Systems, Sandton, South Africa). Over 3 months, 23 polytrauma patients treated at the Emergency Unit of the Red Cross Children's Hospital in Cape Town, South Africa, were imaged on the STATSCAN. Image quality, diagnostic equivalence and clinical efficiency were compared with a computed radiography (CR) system (Fuji FCR 5000, Fuji Photo Film, Tokyo, Japan). The STATSCAN antero-posterior bodygram correlated well technically and diagnostically with CR, showing 96% of the fractures in the cohort. It allowed superior visualisation of the trachea and main bronchi and imaging was, on average, 13% faster than CR. The STATSCAN could play an important role in paediatric polytrauma. The clinical significance of its superior demonstration of the trachea and main bronchi requires further evaluation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.