Abstract

To test for possible thermal injury and tissue damage caused by magnetic-controlled growing rods (MCGRs) during MRI scans. Three fresh frozen cadavers were utilized. Four MRI scans were performed: baseline, after spinal hardware implantation, and twice after MCGR implantation. Cross connectors were placed at the proximal end and at the distal end of the construct, making a complete circuit hinged at those two points. Three points were identified as potential sites for significant heating: adjacent to the proximal and distal cross connectors and adjacent to the actuators. Data collected included tissue temperatures at baseline (R1), after screw insertion (R2), and twice after rod insertions (R3 and R4). Tissue samples were taken and stained for signs of heat damage. There was a slight change in tissue temperature in the regions next to the implants between baseline and after each scan. Average temperatures (°C) increased by 0.94 (0.16-1.63) between R1 and R2, 1.6 (1.23-1.97) between R2 and R3, and 0.39 (0.03-0.83) between R3 and R4. Subsequent histological analysis revealed no signs of heat induced damage. Recurrent MRI scans of patients with MCGRs may be necessary over the course of treatment. When implanted into human cadaveric tissue, these rods appear to not be a risk to the patient with respect to heating or tissue damage. Further in vivo study is warranted. N/A.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.