Abstract

A micro biosensor, which can be applied to a lab-on-a-chip (LOC), is developed in order to detect biomaterials such as protein or DNA. The biomaterials are detected by the mass micro-balancing technique, which measures the change of the resonant frequency of the sensor structure. The sensor structure consists of a micro cantilever actuated by piezoelectric PZT film. The PZT film is designed to act as both a sensor and an actuator. The geometry of the micro-cantilever is determined so as to maximize the sensitivity of the sensor, and environmental effects such as added mass effect in liquid are also considered in the design procedure. The microcantilever is 100 /spl mu/m in length, 30 /spl mu/m in width and 5 /spl mu/m in thickness, and the PZT film thickness is 2.5 /spl mu/m. The first resonant frequency of the PZT-micro cantilever is 1.2/spl sim/1.3 MHz. Lastly, the fabricated micro biosensor using the self-exciting PZT-micro cantilever is tested by detecting the insulin-anti insulin binding protein and poly T-sequence DNA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.