Abstract

A non-invasive piezo-film-based measurement method for haemodynamic assessment is proposed. The design of a system, able to reconstruct the blood pressure waveform online by dealing with problems arising from the piezo-film capacitive nature in the targeted frequency range (from quasi-dc up to 12 Hz), is illustrated. The system is based on a commercial piezo-film placed easily on the radial artery with a special brace without any discomfort for the patient. The analogical conditioning circuit and digital signal processing are continuously tuned with the signal from the sensor to reconstruct the blood pressure signal online. Diagnostic schema, based on physio-pathological models, have been implemented in order to compute online trends of max[dP(t)/d(t)] and volemic status highly useful for the intensivist and anaesthesiologist. The system was characterized by numerical simulation and experimental in vivo comparison to the traditional reference system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.