Abstract

We have constructed an electronically controlled positive-pressure ventilator actuated by phrenic neural activity for use in open-chested or paralyzed experimental animals for the study of breathing pattern. A Bird Mark 14 positive-pressure ventilator was modified such that flow is a linear function of a command signal. Flow is delivered by advancing an air valve with a servo-motor that is controlled by one of three different operational modes. In two of the modes, the difference between the electronic average of inspiratory phrenic activity (moving average) and a feedback signal determines the inspiratory flow. The feedback signal is derived from either tracheal pressure or an electronic measure of inspired volume. In the third mode, the moving average is differentiated to provide control of inspiratory flow and volume. Physiological flow profiles were created using all three operational modes. Integration of an air-valve position signal provides an electronic measure of tidal volume. An additional feature of this ventilator allows inspiratory flow and duration to be predetermined for a given breath.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.