Abstract

The highly contagious norovirus (NoV) is the most common causative agent of acute gastroenteritis, resulting in >200,000 deaths worldwide annually. A rapid and sensitive detection method is a prerequisite for effective prevention and timely identification of NoV contamination. In the present study, we developed a photoelectrochemical (PEC) biosensor coupled with a novel custom-made monoclonal antibody (mAb) for specific and sensitive NoV detection. Our system could detect levels of recombinant NoV capsid protein VP1 as low as 2 × 10−10 g mL−1 (4.9 pM) within 30 min in a concentration-dependent manner. More importantly, the biosensor was versatile in detecting virus isolated from real samples that were as low as 46 copies μL−1. These findings indicate that this system has the potential to serve as a convenient point-of-care system for diagnosing NoV infection and detecting NoV-contaminated food samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.