Abstract

The stent implantation may alter the post-operative patient's blood pressure, and bioresorbable vascular stents (BVS) as a candidate to treat vascular diseases, its degradation is affected by mechanical stress, thus, the altered pressure representing varying stress level will result in different degradation behaviors of the BVS. This paper first proposed a novel stress-regulated PLA degradation model that included swelling factor, and then the degradation evolutions of a PLA BVS within 180days under normal and high blood pressures were simulated by finite element method, and more four degradation indexes were defined to study the effects of the two blood pressures on the degradation of the PLA BVS. The results showed that the high pressure weakly accelerated the degradation of the PLA BVS with respect to the normal pressure by examining the four indexes, e.g., the residual stent volume decreased to 0.72 and 0.69, respectively for the normal and high pressures at day 180. The current finding provided a theoretical understanding of the PLA BVS degradation, and hinted that the PLA BVS may not need to be elaborately selected in clinical practices for treating hypertensive patients.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.