Abstract

PurposeWe present a patient-specific model to estimate tumor location in the thorax during radiation therapy using chest surface displacement as the surrogate signal. MethodsTwo types of data are used for model construction: Four-dimensional computed tomography (4D-CT) images of the patient and the displacement of two points on the patient’s skin on the thoracic area. Principal component analysis is used to fit the correspondence model. This model incorporates the recorded surrogate signals during radiation delivery as input and delivers the 3D trajectory of the tumor as output. We evaluated the accuracy of the proposed model on a respiratory phantom and five lung cancer patients. ResultsFor the respiratory phantom, the location of the center of the sphere during treatment was calculated in three directions: Left-Right (LR), Anterior-Posterior (AP) and, Superior-Inferior (SI). The error of localization was less than 1 mm in the LR and AP directions and less than 2 mm in the SI direction. The location of the tumor center for two of the patients, and the location of the apex of the diaphragm for the other three, was calculated in three directions. For all patients, the localization error in the LR and AP directions was less than 1.1 mm for two fractions and the maximum localization error in the SI direction was 6.4 mm. ConclusionsThis work presents a feasibility study of utilizing surface displacement data to locate the tumor in the thorax during radiation treatment. Future work will validate the model on a larger patient population.